[Translate to fr:] Der medizinische 70 Watt Auto-Adapter CAR70M mit Car/Truck Eingangsstecker.

CAR70M

Adaptateur de voiture médical 70W

  • Large plage de tension d'entrée de 10-32V
  • Puissance de sortie maximale de 70W
  • Isolation galvanique pour une sécurité maximale du patient
  • Connecteur d'entrée voiture/camion
  • Fixation murale en option

Vous souhaitez obtenir plus d'informations ?

Downloads CAR70M

Fiche produit CAR70M

Specification

Input
Voltage range10.00 - 32.00VDC
Standby currentTypical 37.00mA @ 10.00V input
Typical 29.00mA @ 32.00V input
Input fuse10.00A
ProtectionReverse polarity (input fuse),
Under voltage protection (Transient suppression circuit)
Output
Voltage range16.00 - 19.00V nominal
Power

66.00W cont.
70.00W max.

Initial set tolerance

±3% max.

Load regulation

-4% max.

Line regulation

±1% max.

Ripple & noise

< 0.12V pk-pk

Current limitConstant current
Protection

Short circuit
Over temperature protection
Over voltage protection

Environmental

CoolingConvection cooled
Temperature

Operating: -20°C to 40°C (40°C to 50°C, derating 35%)
Non-operating: -40°C to 75°C

Pressure & altitude

Operating: 1200hPa to 570hPa (max. 4600m above sea level)
Non-operating: 1200hPa to 115hPa (max. 15000m above sea level)

Humidity5 to 95% r.H., non condensing
General
Galvanic isolationIsolation voltage 2.00kV
Input connector

Cigarette lighter adapter

Output connector

Customer-specific

Efficiency

typical 88% at 100% load

MTBF

> 100000h @ 25°C ambient temperature and Gper MIL-HDBK217F Humidity

Green procurementRoHS 2011/65/EU + 2015/863/EU
REACH 1907/2006
WEEE 2012/19/EU
IndicatorSingle green LED
Ingress protection rating

IEC60529, IP42

Regulatory approvalsEuropeCE
 United KingdomUKCA
 International

CB Report IEC 60601-1 2nd edition
CB Report IEC 60601-1 3rd edition

 USA & CanadaUL60601-1
Electromagnetic emissionsEurope

EN 55011/55022 class B

 USA & CanadaFCC15 class B
 Medical electrical equipment

IEC60601-1-2

Electromagnetic immunityESD immunity

IEC61000-4-2, 8kV contact/15kV air

 Electromagnetic field immunity

IEC61000-4-3, 80-2700MHz, 30V/m, modulation 80%/200Hz

 Electrical fast transient/burst immunity test

IEC61000-4-4, +/-2kV, 5/50ns

 Surge immunity test

IEC61000-4-5, +/-0.5kVsym, 1.2/50(8/20)μs

 Conducted Immunity

IEC61000-4-6, 0.15-80MHz, 10V, 80%, 1KHz, AM, with CDN

 ISO-Impulses

ISO7637-2, pulse 1 till 5, immunity test level III

 Medical electrical equipment

IEC60601-1-2

 UNECE vehicle regulationE13 ECE Reg. 10

- Other approvals on request

Housing dimensions (LxWxH)

160.00 x 59.00 x 34.00mm, without cables and connectors 

Input cable length

240cm ±5%

Output cable length

100cm ±5%

Weight

approx. 350g, without cables, connectors and wall-mount kit


CAR70M est homologué dans le monde entier

United Kingdom